▶ Small and easy portable
▶ Features
Profile | Profile infusion condition (flow rate, delivery volume, delivery time) differently every hour, up to 24 hours. |
---|---|
Drug labeling (EASY MODE) | Save 3 of drug's names and infusion data |
Self testing | Every time the unit turns on, self testing is proceeded. |
K.V.O.(Keep Vein Open) | When the delivery volume has been reached, flow rate turns into K.V.O. rate ( adjustable from 0.1ml/h ~0.9ml/h) automatically. |
Keypad lock function | Keypad is locked with or without password depend on its using condition. |
Infusion setup | Flow rate / Delivery volume / Total infused volume |
Buzzer | 3 step |
Occlusion level | 3 step (4psi~20psi) |
Bolus | ON/OFF, Adjustable (1~30ml/h) |
History call back | Display up to last 15 infusion data |
Data saving years | 5 years |
Strengthen safety | Double CPU stop on error |
Avoiding FREE FLOW | |
Dosage mode(Body weight mode) | Automatically calculate |
Recharge battery | |
Battery recharging time | 4 hours/td> |
Battery operating time | 4 hours (25ml/h ) |
▶ Specification
FLOW RATE | 0.1 ~ 250ml/h |
---|---|
BOLUS RATE | 0.1~60.0ml/h |
DELIVERY VLUME | 0.1ml ~ 999.9ml (in 0.1ml increments) |
PURGE RATE | 1mL/h ~ 30.0mL/h |
KVO RATE | 0.1mL/h ~ 0.9mL/h |
LOCK TIME | 1~ 999min (±1min) |
ACCURACY | ±3% |
OPERATING METHOD | Peristalsis |
LEVEL OF OCCULATION | 4psi ~ 20psi ( 3 STEPS ) |
PROFILE FUCTION | Save event, Language selection (optional) Connect to USB , History call back |
POWOR | A/C : 100-240 V~ / 50-60Hz, DC 7.4V Class 1, Internal power supply, TypeCF |
OPERATION CONDITION | +5 ~ 45°C, 90% RH (no condensation) |
STORAGE CONDITION | -20 ~ 60°C, 95% RH (no condensation) |
ALARM | AIR IN LINE, OCCLUSION, DOOR OPEN, EMPTY CONTAINER, K.V.O. RATE (INFUSION COMPLETION),LOW BATTERY, REPEAT ALARM, FLOW ERROR (I.V. SET FREE FLOW PROTECTION), DEIVCE MALFUNCTION |