LOGO

Micro Digital

Biotechnology, Optical measuring device and Automation system manufacturer
Name

Fully Automatic Immuno Analyzer

Model Diamond
Series Products
Product Description

Diamond

Fully Automatic Immuno Analyzer 

Diamond – Fully Automatic Immuno-Analyzer takes care of sample and reagent handling to well washing and optical measurement with a touch of a button.

Diamond is a convenient and accurate fully automated system specialized for CLIA(chemiluminescence immunoassay) and ELISA(enzyme-linked immunosorbent assay)

 

Feature

 - Fully automated immunoassay system (luminescence, absorbance measurement)
 - Open platform
 - Compatible with conventional 96-well plates
 - Quick and accurate control
 - Minimized inter-well cross contamination through the use of disposable tips
 - Accurate liquid handling

 

Micro Digital Fully Automatic Immuno Analyzer Diamond

 

Specification

 

General
Sample Capacity 96 patient sample positions
Reagent Capacity 9 custom reagent positions
Test Tube Standard 13, 16mm diameter (or customizable)
No. of Microplates 2
Well-Tracking Independent well tracking & timing
Barcode Reader Optional
System Architecture Universal open platform (Qualitative, Quantitative)
Sample Tray Easy loading and unloading of samples
Dimension 650x830x650mm (WxDxH)
Power 100~240V, 50~60Hz
ELISA Reader specification
Dynamic Range 0~3.0 OD
Photometer 450, 630nm (405, 490, 550 optional)
Filter Slots 6
Precision <1%CV (2.0 OD)
Accuracy ±0.005 OD or 2.5%
CLIA Reader specification
Dynamic Range Up to 7-log of dynamic range
Limit of Detection 1x10-18 moles of ATP
Crosstalk 3.0x10-5
Washer
Manifold Configuration 8 channel
Dispense-volume Range 100~300ul
Washing Tank 2 external washing tank and 1 waste tank
Wash Cycle 1~∞ (customizable)
Residual Volume <1ul
Incubator
Temperature Range Up to 42℃
Temperature Uniformity ±0.5℃
Shaking Amplitude : 1mm, speed : up to 800rpm
Temperature Monitoring Yes
Pipetting
Aspiration Volume >25ul
Tip Disposable
Serial Dilution Yes
Replicates Yes
Precision <2% CV at 100ul
Process Security
Liquid-level Sensing Yes (Only provided disposable tips from MicroDigital., Ltd)
Level-sensor System Pressure differential
Clot detection Yes (Optional)
Dispense-anomaly Detection Yes (Optional)
Tip detection Yes (Optional)
Well-fill Verification Yes
Alarms Warning message on PC software
Software
Operating Software 3 types (Developer, User, Hospital mode)
Work Protocols (Assays) Unlimited
Data Processing Quantitative and Qualitative
Access Level Control Yes (Master, Manager, Operator)
O/S MS window XP, 7, 10