Diamond
Fully Automatic Immuno Analyzer
Diamond – Fully Automatic Immuno-Analyzer takes care of sample and reagent handling to well washing and optical measurement with a touch of a button.
Diamond is a convenient and accurate fully automated system specialized for CLIA(chemiluminescence immunoassay) and ELISA(enzyme-linked immunosorbent assay)
Feature
- Fully automated immunoassay system (luminescence, absorbance measurement)
- Open platform
- Compatible with conventional 96-well plates
- Quick and accurate control
- Minimized inter-well cross contamination through the use of disposable tips
- Accurate liquid handling
Specification
General | |
Sample Capacity | 96 patient sample positions |
Reagent Capacity | 9 custom reagent positions |
Test Tube | Standard 13, 16mm diameter (or customizable) |
No. of Microplates | 2 |
Well-Tracking | Independent well tracking & timing |
Barcode Reader | Optional |
System Architecture | Universal open platform (Qualitative, Quantitative) |
Sample Tray | Easy loading and unloading of samples |
Dimension | 650x830x650mm (WxDxH) |
Power | 100~240V, 50~60Hz |
ELISA Reader specification | |
Dynamic Range | 0~3.0 OD |
Photometer | 450, 630nm (405, 490, 550 optional) |
Filter Slots | 6 |
Precision | <1%CV (2.0 OD) |
Accuracy | ±0.005 OD or 2.5% |
CLIA Reader specification | |
Dynamic Range | Up to 7-log of dynamic range |
Limit of Detection | 1x10-18 moles of ATP |
Crosstalk | 3.0x10-5 |
Washer | |
Manifold Configuration | 8 channel |
Dispense-volume Range | 100~300ul |
Washing Tank | 2 external washing tank and 1 waste tank |
Wash Cycle | 1~∞ (customizable) |
Residual Volume | <1ul |
Incubator | |
Temperature Range | Up to 42℃ |
Temperature Uniformity | ±0.5℃ |
Shaking | Amplitude : 1mm, speed : up to 800rpm |
Temperature Monitoring | Yes |
Pipetting | |
Aspiration Volume | >25ul |
Tip | Disposable |
Serial Dilution | Yes |
Replicates | Yes |
Precision | <2% CV at 100ul |
Process Security | |
Liquid-level Sensing | Yes (Only provided disposable tips from MicroDigital., Ltd) |
Level-sensor System | Pressure differential |
Clot detection | Yes (Optional) |
Dispense-anomaly Detection | Yes (Optional) |
Tip detection | Yes (Optional) |
Well-fill Verification | Yes |
Alarms | Warning message on PC software |
Software | |
Operating Software | 3 types (Developer, User, Hospital mode) |
Work Protocols (Assays) | Unlimited |
Data Processing | Quantitative and Qualitative |
Access Level Control | Yes (Master, Manager, Operator) |
O/S | MS window XP, 7, 10 |